Detail Produk

Operational Life :

When radiant energy reaches 50% of the initial energy at 300mA(dc).

 

Warranty:

  1. Do not subject the lamp to mechanical shock.
  2. Before using the lamp, ensure glass envelope is clean and free from dirt. Envelope may be gently wiped with either alcohol or acetone.
    Caution: the circumference of the glass envelope around the light box is very thin and should be handled with extreme care.
  3. Care must be taken not to touch lamp body with bare hand, as it’s temperature reaches over 200°C during operation.
  4. UV radiation is harmful to eyes. Suitable eye protection must be worn when observing light output.
  5. Since this lamp operates at high voltages, extreme caution is require to prevent electrical shock.

 

 

TYPE : P-714

D2 Lamps – (Deuterium Lamps)

 

Aplication :

  • UV-Visible Spectroscopty,
  • Spectrofluorimetry
  • UV detectors attached to HPLC instruments,
  • Colorimetry,
  • Refractometry and as a non-atomic source for background correction in Atomic Absorbtion Spectroscopy.

 

INSTRUMENT :

Spectra Physics

ELECTRICAL :

 

Filament / Heater VoltageStart2.5 – 3.0 VDC
Operating1.5 – 2.0
Filament / Heater Current4.5A Max
Discharge / Arc VoltageStrike@25°C >350 VDC
Operating60 – 90 VDC
Discharge / Arc Current10 -300mA
LeadsBlackFilament & Cathode
RedAnode
Terminals3 Pin housing

 

OPTICAL :

Window materialUV Transmitting Borosilicate
Source apertureØ 1mm Diameter
Wavelength range185nm – 400nm
Noise<1 x 10-4 Absorption units
Drift< 3 x 10-3 Absorption units